Reactivity Test Apparatus
Ground quicklime shall be tested for their reactivity with this apparatus which consists essentially of a Dewar vessel, 1000 ml cap., stirring motor, calibrated thermometer, stand and accessories.
Two versions are available:
- 64-L0035/D: with advanced datalog thermometer, temperature probe, serial cable for PC connection and dedicated software for download data
- 64-L0035/E: supplied with digital thermometer
Dimensions, approx.: 350x300x800 mm
Weight approx.: 9.5 kg
Two versions available:
- with standard digital thermometer (64-L0035/D)
- with datalog thermometer complete with temperature probe (64-L0035/E)
64-L0035/D – datalog version:
Apparatus for testing reactivity of quicklime coupled with datalog thermometer (64-L0035/D1):
- max temperature: 900°C
- resolution: 0.1°C (from -149.9 to 900°C); 0.2°C (from -200.0 to -150.0°C)
- accuracy: ±0.5°C (form -100.0 to 900°C); ±1°C (over);
- °C to °F shifting
- environment: -10 to 50°C (14 to 122°F); RH max 100%
- PC connection: RS232 serial port
- battery operated
- 110-230 V, 50-60 Hz, 1 ph.
64-L0035/E – digital version:
Apparatus for testing reactivity of quicklime coupled with digital thermometer (64-L0035/E1):
- Max.temperature: 220° C
- Resolution: 0.1° C
- Accuracy: ± 0.3° C
- 110-230 V, 50-60 Hz, 1 ph.
Code
Description
64-L0035/D
Apparatus for testing reactivity of quicklime to EN459-2. Supplied with digital thermometer, serialcable for PC connection and software for data download. 110-240V/50-60Hz/1Ph.
64-L0035/E
Apparatus for testing reactivity of quicklime to EN459-2. Supplied with digital thermometer. 110-240V/50-60Hz/1Ph.
Code
Description
Spare Parts
64-L0035/C1
Spare Dewar vessel for 64-L0035/x apparatus supplied without stopper
64-L0035/C2
Stirring paddle for 64-L0035/x
64-L0035/D1
Spare thermometer for 64-L0035/D
-Max. temperature: 900°C
-Resolution: 0.1°C
-Accuracy: +/- 0.5°C
64-L0035/E1
Spare thermometer for 64-L0035/E.
-Max.temperature: 220° C
-Resolution: 0.1° C
-Accuracy: ± 0.3° C