Filterless Centrifuge Binder Extractors are used for rapid filterless separation of filler (ash) from binder solution coming, for instance, from wire mesh extractor model 75-B0015.

Enquire Now

The centrifuge can also be used for binder recovery from an asphalt sample previously disintegrated  in solvent and poured into the funnel fitted with 200 mm dia. test sieves to separate gradually the aggregates.

How is the test carried out

The test is carried out by pouring the solvent from the top funnel into the rotating aluminium cup.

Due to the centrifugal effect, the liquid spreads on the wall and moves upwards, leaving mineral particles in the cup whilst the liquid is discharged outside through the outlet tubing.

A special electronic circuit provides for a preset controlled ramp which gradually increases the speed to the maximum.

Two models available:

  • 75-B0024/P for filler extraction up to 100 g  per test, fit with 70 mm dia. cup
  • 75-B0024/B for filler extraction up to 400 g per test, fit with 122 mm dia. cup.

The 75-B0024/P, PY and PZ versions do not include sieves which have to be ordered separately, conforming to EN standard. See accessories. The cup should be internally lined with paper for a better and complete removal of filler. See accessories: 75-B0005/2.

The solvent used can be profitably recovered with the  solvent recovery unit.

Models75-B0024/P
75-B0024/PY
75-B0024/PZ
75-B0024/B
Max. quantity of filler extracted per test50-100 g400 g
Cup dimensions, mmDia. 70×190Dia. 122×211
Max. speed, r.p.m.1100011000
Power, W5501000
Sieves includedSee accessories0.075, 0.25,
0.71, 2 mm
Overall dimensions (lxwxh) mm500 x 370 x 850560 x 640 x 1200
Weight approx. kg:55100

Code
Description
75-B0024/B
Continuous flow filterless centrifuge 122 mm dia. cup. 380V, 50Hz, 3ph.
75-B0024/P
Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 230V/50Hz/1Ph.
75-B0024/PY
Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 220V/60Hz/1Ph.
75-B0024/PZ
Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 110V/60Hz/1Ph.
Code
Description

Any question about the Filterless Centrifuge Binder Extractors

"*" indicates required fields

Controls Brand Logos

Filterless Centrifuge Binder Extractors

Filterless Centrifuge Binder Extractors are used for rapid filterless separation of filler (ash) from binder solution coming, for instance, from wire mesh extractor model 75-B0015.

Features

Specifications

Models75-B0024/P
75-B0024/PY
75-B0024/PZ
75-B0024/B
Max. quantity of filler extracted per test50-100 g400 g
Cup dimensions, mmDia. 70×190Dia. 122×211
Max. speed, r.p.m.1100011000
Power, W5501000
Sieves includedSee accessories0.075, 0.25,
0.71, 2 mm
Overall dimensions (lxwxh) mm500 x 370 x 850560 x 640 x 1200
Weight approx. kg:55100

Highlights

The centrifuge can also be used for binder recovery from an asphalt sample previously disintegrated  in solvent and poured into the funnel fitted with 200 mm dia. test sieves to separate gradually the aggregates.

How is the test carried out

The test is carried out by pouring the solvent from the top funnel into the rotating aluminium cup.

Due to the centrifugal effect, the liquid spreads on the wall and moves upwards, leaving mineral particles in the cup whilst the liquid is discharged outside through the outlet tubing.

A special electronic circuit provides for a preset controlled ramp which gradually increases the speed to the maximum.

Two models available:

  • 75-B0024/P for filler extraction up to 100 g  per test, fit with 70 mm dia. cup
  • 75-B0024/B for filler extraction up to 400 g per test, fit with 122 mm dia. cup.

The 75-B0024/P, PY and PZ versions do not include sieves which have to be ordered separately, conforming to EN standard. See accessories. The cup should be internally lined with paper for a better and complete removal of filler. See accessories: 75-B0005/2.

The solvent used can be profitably recovered with the  solvent recovery unit.

Models

  • 75-B0024/B/ Continuous flow filterless centrifuge 122 mm dia. cup. 380V, 50Hz, 3ph.
  • 75-B0024/P/ Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 230V/50Hz/1Ph.
  • 75-B0024/PY/ Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 220V/60Hz/1Ph.
  • 75-B0024/PZ/ Continuous flow filterless centrifuge 70mm dia. cup. Sieves not included. 110V/60Hz/1Ph.

Accessories

Sieves
  • 15-D2300/J/ 200 mm dia ISO (BS) sieve op.150 microns
  • 15-D2335/J/ 200 mm dia ISO (BS,UNI,NF,DIN) sieve op.63 microns
  • 15-D2185/J/ 200 mm dia ISO (BS,UNI,NF,DIN) sieve op.2,00 mm
  • 15-D2230/J/ 200 mm dia ISO (BS,UNI,NF,DIN)sieve op.710 microns
  • 15-D2275/J/ 200 mm dia ISO (BS,UNI,NF,DIN)sieve op.250 microns
  • 15-D2330/J/ 200 mm dia ISO (BS,UNI) sieve op.75 microns
Spare Cup and Line Paper
  • 75-B0005/1/ Spare cup for centrifuge
  • 75-B0005/2/ Line paper for centrifuge cup. Pack of 100 pieces.
Aluminium Beaker
  • 75-B0024/1/ Aluminium beaker

Standards

  • ASTM D2172
  • EN 12697-1