What are the main advantages of submersible load cells in comparison to external load cells in a triaxial system?
Submersible load cells have been designed to work inside triaxial cells. They have a lower hysteresis and very good linearity as well as coming with a substantial overload safety feature.
Advantages:
- Load reading remains unaffected throughout pressure range of use.
- No need to compensate for ram up thrust
No need to compensate for ram friction
What system do I need for performing a triaxial stress path test?
Stress path test requires a PC-controlled triaxial system with close loop load control such as an Autotriax EmS or Dynatriax EmS fitted with the following mandatory accessories:
- Banded triaxial cell with vacuum attachment (e.g. 28-WF4070, 28-WF4100, 28-WF4150) for connecting vacuum top cap
- Vacuum top cap to replace the standard top cap, ensuring that the submersible load cell and sample are in contact with each other when closing the triaxial cell
- Submersible load cell to eliminate ram friction
- Specific management test software with user-defined independent ramps/hold deviator stress, cell pressure and back pressure.
Can I overload my machine?
NO – All CONTROLS machines have 2 protection systems preventing overloading
Can I overextend ram travel?
NO – All CONTROLS machines come with ram travel limit switch installed as a standard outfit